6.1 REFERENCE CONDITIONS

Parameter
Ambient temperature
Relative humidity
Voltage
Current
Phase voltage
Input voltage of current inputs
(except AmpFlex
®
Network frequency
Harmonics
Preheating
Common mode
Magnetic field
Electrical field

6.2 ELECTRICAL PROPERTIES

Uncertainties are expressed in % of reading (R) and an offset: ± (a % R + b).

6.2.1 Voltage Inputs

Operating range: Up to 600 V
NOTE: Phase-to-neutral voltages less than 2 V and 1200 V
phase-to-phase voltages less than 3.4 V are set to zero.
Input impedance: 903 K when the device is running on battery.
Sustained overload: 660 V. Above 690 V, the device will display the OL
46

6. SPECIFICATIONS

Reference conditions
(73.4 ± 3.6) °F (23 ± 2) °C
(45 to 75) % RH
No DC component.
No DC component.
[100 V
[50 mV; 1.2 V] without DC (< 0.5 %)
/MiniFlex
)
®
(50 ± 0.1) Hz and (60 ± 0.1) Hz
< 0.1 %
The device must have been powered up for at
least one hour.
The neutral input and the casing are grounded.
The device is powered by the battery.
0 A/m AC
0 V/m AC
RMS
voltages from (45 to 65) Hz.
When the device is powered by the voltage at the terminals,
the impedance on L1 is dynamic and the current source must
be able to deliver up to 100 mA at 90 V and 500 mA at 600 V.
symbol.
Power & Energy Logger Model PEL 52 - User Manual
; 600 V
] without DC (< 0.5 %)
RMS
RMS
Table 3
for phase-neutral and phase-phase
for
RMS