Preparing Forceps Valve And Button; Method Of Cleaning, Disinfecting And Sterilizing Forceps Valve - FujiFilm EN-580T Operation Manual

Endoscope

Advertisement

Chapter 5 Preparation for Use of Endoscope
5.2

Preparing Forceps Valve and Button

5.2.1

Method of Cleaning, Disinfecting and Sterilizing Forceps Valve

(1) Wipe the surface of the parts with gauze soaked in
cleaning fluid.
(2)
Brush the entire forceps valve in the cleaning fluid using a
soft toothbrush.
(3) Clean the forceps valve using the ultrasonic cleaning unit.
(4)
Pull out the forceps valve from the cleaning fluid and rinse
off with running water (clean water).
(5)
Wipe off the remaining water on the parts with dry sterile
gauze.
5-4
For operational methods of the ultrasonic cleaning unit/autoclaving unit, follow the
instructions described in each operation manual for the respective units.
[Note] Do not use the forceps valve after its expiration date has elapsed.
The forceps valve is a single-use product. Prepare a new forceps valve before use.
The forceps valve has not been sterilized. Perform cleaning and disinfection or sterilization
before use according to the following procedure.