Inspecting of the
red laser beam
step-by-step
Builder, Check & Adjust
1. Set up the provided CPR105 Flat prism between 5 m and 20 m with the reflective
tape side facing the instrument.
2. Align the instrument crosshairs to the centre of the Flat prism.
3. Switch on the red laser beam by activating the Laserpointer function in Config-
uration Mode (except Builder 400).
4. Without using the telescope inspect the position of the red laser dot on the Flat
prism.
Look at the Flat prism from just above the telescope or from just to the side of it.
5. If the dot is within the inner printed circle the laser beam is within tolerance. If
it is outside it is recommended to have the laser beam realigned by a Leica
Geosystems authorized service workshop.
193