Performance Test Instructions

Agilent 8163A/B Mainframe with
fixed laser source and power sensor
Laser
Source
Agilent 8166A/B Mainframe with fixed laser source and power sensor
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Performance Test Instructions
Parameters to be tested:
• Backplane Connectors by Module Performance Test
• Power Supply by Module Interaction Test
• Display by Keystroke Test
Power
Sensor
Figure 199
Mainframe Setup

Display/Key Functional Test

1 Setup the test equipment as shown in Figure 199:
Insert the fixed laser source module in slot 1 and the power sensor
in slot 2. If you use the Agilent 8164A/B Lightwave Measurement
System, insert the back-loadable Tunable Laser module, if available,
in slot 0.
Agilent 8164A/B Mainframe with fixed laser source,
power sensor, and tunable laser module
Performance Tests
Laser
Power
Source
Sensor
Tunable Laser
341