Cleaning Instructions

What do I need for proper cleaning?

Furthermore, the power density may burn dust into the fiber and
cause additional damage (for example, 0 dBm optical power in a single
2
mode fiber causes a power density of approximately 16 million W/m
).
If this happens, measurements become inaccurate and non-repeatable.
Cleaning is, therefore, an essential yet difficult task. Unfortunately,
when comparing most published cleaning recommendations, you will
discover that they contain several inconsistencies. In this section, we
want to suggest ways to help you clean your various optical devices,
and thus significantly improve the accuracy and repeatability of your
lightwave measurements.
What do I need for proper
cleaning?
Some Standard Cleaning Equipment is necessary for cleaning your
instrument. For certain cleaning procedures, you may also require
certain Additional Cleaning Equipment.

Standard Cleaning Equipment

Before you can start your cleaning procedure you need the following
standard equipment:

• Dust and shutter caps

• Isopropyl alcohol
• Cotton swabs
• Soft tissues
• Pipe cleaner
• Compressed air
Dust and shutter caps
All of Agilent Technologies' lightwave instruments are delivered with
either laser shutter caps or dust caps on the lightwave adapter. Any
cables come with covers to protect the cable ends from damage or
contamination.
412
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition