Power Measurement
How to Measure Power
10 Move to the [Range Mode] parameter and press Enter.
11 Move to <Manual>, by using the cursor key, and press Enter.
12 Move to the [Range] parameter and press Enter.
13 Enter the range value determined in Step 8, by using the cursor key,
and press Enter.
Figure 48
Manual Range Mode - Within Range
If the measured power is more than 3dBm greater than the range
setting, it is impossible for power to be displayed. The power value,
+1.--- dBm, as shown in Figure 49, is shown. This means that the
measured power is greater than the Upper Power Limit. You must
decrease the [Range] value.
90
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition