Error Messages
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
398
E10080: Laser <a> control for safety test failed
To test the current safety comparator the laser current needs to
be set to two different values
a
Please return the unit to your local service center for repair
E10081: Safety comparator <a> test failed (b, c)
Checks whether laser current safety comparator includes at the
right conditions a safety interrupt at the CPU
a
laser number
b
shows whether the interrupt flag was raised when the
current was lower than the limit
c
shows whether the interrupt flag was raised when the
current was higher than the limit
Please return the unit to your local service center for repair
E10090: Peltier <a> regulation failed (b, c, d)
During boot time, when the Peltier control was disabled it is en-
abled for two seconds and the temperature moving direction is
checked. If the movement delta is not approaching the target de-
fault temperature or if the temperature has not reached the de-
fault within the given ranges this test fails
a
laser number
b
default temperature
c
temperature with disabled peltier
d
temperature with enabled peltier
Please return the unit to your local service center for repair
E10091: Peltier <a> regulation failed (b, c)
Verifies whether the temperature is around the default value with
the given limits
a
laser number
b
default temperature
c
actual temperature
Please return the unit to your local service center for repair