E10081: Safety comparator <a> test failed (b, c)
Checks whether laser current safety comparator includes at the
right conditions a safety interrupt at the CPU
a
laser number
b
shows whether the interrupt flag was raised when the
current was lower than the limit
c
shows whether the interrupt flag was raised when the
current was higher than the limit
Please return the unit to your local service center for repair