Powermeters and Interface Modules with Optical Head

Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Powermeters and Interface
Modules with Optical Head
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
E10010: Invalid Caldata
The Caldata loaded on the module is invalid. Either the caldata
version number is not compatible with the firmware or the calda-
ta is corrupt (checksum incorrect)
(no test value)
Please return the unit to your local service center for repair
E10012: Dual ported RAM write/read test failed
Writing bit patterns and reading them back to/from the dual port-
ed ram failed
(no test value)
Please return the unit to your local service center for repair
E10013: FPGA not correctly initialized
Loading the FPGA with data failed. This indicates a defective
FPGA or an FPGA access problem
(no test value)
Please return the unit to your local service center for repair
E1x020: Communication with head failed
An external head is available but it was not possible to access
the registers of the head's FPGA
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x021: Invalid head caldata
The head calibration data was detected as invalid
x
refers to the module channel tested
Please return the unit to your local service center for repair
Error Messages
391