378
Error
Error
Channel <slot>.<channel>
Sweep Parameter Inconsistent
One or more of the following reasons may cause this error message:
all
stepped sweep
lambda logging on
switch lambda logging off
<= 2.5x
continuous sweep
stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- cycles ≠ 1
- trigger ≠ step finished
check settings and correct where appropriate
> 2.6
continuous sweep
stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- trigger ≠ step finished
check settings and correct where appropriate
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Errors appearing on pop-up menus