How To Use Output Triggering - Agilent Technologies 8163A User Manual

Lightwave multimeter,lightwave measurement system,lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

How to Use Laser Source Modules
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition

How to Use Output Triggering

You can configure the output trigger connector to output a TTL signal
at the frequency of the internal modulation of a Agilent 8163A/B
Series Laser Source module.
To set your module's Output Trigger Configuration:
1 See "How to Set the Trigger Configuration" on page 65 for how to
configure the trigger connectors.
2 Move to the Laser Source channel and press [Menu].
3 Move to <Output Trigger Mode>, by using the cursor key, and press
Enter. You will see the screen in Figure 62.
Figure 62
Output Trigger Mode
4 Move to one of the following, by using the cursor key:
– <Disabled>, the output trigger mode is disabled.
– <Modulation>, the output trigger connector outputs a TTL signal
at the frequency of the internal modulation. This signal is output
whether the laser is switched on or off.
5 Press Enter.
Laser Sources
111

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8166a8166b8163b8164a8164b

Table of Contents