Error Messages
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
392
E1x031: 0.25V ref out of limits (a, b, c)
The 0.25V reference voltage measured through the 16bit ADC is
out of the acceptable limits
a
shows the ADC reading (0.25 = 3276)
b
shows the lower limit
c
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
E10040: 10bit ADC test failed (a, b, c)
The 10bit ADC has returned a value other than 512
a
shows the ADC reading
b
shows the lower limit
c
shows the upper limit
Please return the unit to your local service center for repair
E1x050: TEC NTC temp ou of limits (a, b, c, d)
The sensor temperature is out of the acceptable limits
a
shows the sensor NTC temperature
b
shows the sensor NTC temperature ADC reading
c
shows the lower limit
d
shows the upper limit
x
refers to the module channel tested
Verify whether the module is operated within the specified ambi-
ent temperature. If the failure persists, please return the unit to
your local service center for repair.
E1x051: TEC current out of limits (a, b, c)
Verifies whether the TEC current is within a given range
a
shows the ADC reading of the TEC current
b
shows the lower limit
c
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Powermeters and Interface Modules with Optical Head