Error Messages
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
384
TCFS Temp. Selftest Failed (error value)
Verifies if the peltier temperature is within target with a given uncertainty
15010
TCFS peltier temperature is not in target range
Please return the unit to your local service center for repair
TCFS Resistor Selftest Failed (error value)
This test measures the R
15021
Photodiode R
15022
Photodiode R
15023
Photodiode R
Please return the unit to your local service center for repair
TCFS Attenuator Selftest Failed (error value)
Verifies whether an attenuator position is repeatable.
15030
Attenuator could not be positioned
Please return the unit to your local service center for repair
Main DC Selftest Failed (error value)
Tests all fixed DC voltages available at the mainboard mux
23011
+15V Measurement DC voltage not in limits
23012
-15V Measurement DC voltage not in limits
23013
+5V Reference DC voltage not in limits
23014
-5V Reference DC voltage not in limits
23015
negative supply voltage of laser current source not in limits
23016
positive pezo supply voltage not in limits
Please return the unit to your local service center for repair
Main Temperature Selftest Failed (error value)
Checks whether the temperature sensors are within given limits
23021
internal temperature sensor above limit
23022
T_FIRE_DIODE is possibly not connected
23023
T_OPTIZ_DIODE is possibly not connected
23024
T_BASEPLATE is possibly not connected
Please return the unit to your local service center for repair
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
of the photodiodes at channel A, B and M
d
in channel A is below limit
d
in channel B is below limit
d
in channel C is below limit
d
Tunable Laser Sources