Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
TCFS Shutter Selftest Failed (error value)
This test verifies the shutter driver by applying a designated test pattern
25060
TCFS shutter driver electronic fails
Please return the unit to your local service center for repair
Return Loss Meters
E10010: Invalid Caldata
The Caldata loaded on the module is invalid. Either the caldata ver-
sion number is not compatible with the firmware or the caldata is
corrupt (checksum incorrect)
(no parameter)
Please return the unit to your local service center for repair
E10012: Dual ported RAM write/read test failed
Writing bit patterns and reading them back to/from the dual ported
ram failed
(no test value)
Please return the unit to your local service center for repair
E10013: FPGA not correctly initialized
Loading the FPGA with data failed. This indicates a defective FPGA
or an FPGA access problem
(no test value)
Please return the unit to your local service center for repair
E10014: Firmware corrupt, checksum error (a,b)
Firmware in flash has an invalid checksum. Either the image which
was programmed was corrupt or the flash memory has a failure
a
is the stored checksum
b
is the calculated checksum
Reload the module firmware. If the failure persists, please return the
unit to your local service center for repair
Error Messages
387