Error Messages
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
402
c
shows the upper limit
Please return the unit to your local service center for repair
E1060: 0.25V offset failed (a, b)
Verifies that the offset voltage for the 16-bit ADC (250 mV) is cor-
rectly added to the input signal
a
shows the ADC reading
b
shows the maximum difference to the exact ADC reading
of 3277
Please return the unit to your local service center for repair
E10061: PGA DAC adj failed (a, b, c)
Tries to adjust the offset voltage of the PGA
a
shows the ADC Offset reading
b
shows the limit for an upper offset
c
shows the limit for a lower offset
Please return the unit to your local service center for repair
E10062: PGA gain adj failed (a, b, c)
Verifies the gain settings of the PGA (x 1/10/100). A voltage of 25
mV is applied to the input of the PGA and then amplified
a
if 1, the required division has failed
if 0, the calculated factors are correct
b
shows the measured values for an amplification by 10
c
shows the measured values for an amplification by 100
Please return the unit to your local service center for repair
E10063: Lambda Dac adj failed (a, b, c, d)
Verifies the gain of the responsiveness correction stage
a
if 1, the quotient d could not be calculated, otherwise 0
b
shows the difference between DAC settings 1024 and
2048
c
shows the difference between DAC settings 2048 and
4096
d
shows the quotient of b divided by c
Please return the unit to your local service center for repair
The following error codes are attenuator module specific:
– Error Creating message queue!
– Error Init MCB!
– Error spawning tLowlvlMsgHandler!
– Error spawning tKeyPressed!
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Attenuator Modules