Applications
254
23 When the PACT Reference Measurement completes, you hear a
beep, an hourglass appears momentarily.
If there is another Power Meter channel to reference, you will be
prompted to perform this reference measurement from the user
interface by a prompt box similar to the box in Figure 164. Restart
this procedure at step 21.
If you are viewing the PACT Reference Measurement Finished box,
reference sweeps have been performed for all selected Power
Meters.
Figure 165
PACT Reference Measurement Finished Box

How to Perform a Loss Measurement

24 To return to the Setup screen, pressing [Menu], move to <Setup>,
and press Enter.
N O TE
If you change any measurement parameters, the reference
measurement becomes invalid.
25 Connect the Tunable Laser module to the Power Meter channel
through all the system components, including the Device Under
Test. Press [Measure].
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
The PACT Application