Error Messages
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
380
Initialization Error (DspMainCalcKFactor failed, error value)
DSP K factor calibration failed
-1
no TCFS (according to caldata)
1
error when calibrating TCFS
2
DSP sweep failed
4
laser safety
Please return the unit to your local service center for repair
Initialization Error (DspMainWriteCaldata(), error value)
Error while writing cal data to DSP
5
tried to access invalid array position
6
wrong dimension
7
unknown type found in variable array
8
wrong parameter when trying to access one of the DSPs cal
arrays
Please return the unit to your local service center for repair
Initialization Error (Firmware CRC wrong: crc1, stored: crc2!)
The checksum stored in the firmware does not match the calculated
checksum
crc1
calculated checksum
crc2
checksum read from caldata
Reload the firmware of the module. If the failure persists, please return
the unit to your local service center for repair
Initialization Error (HwAccessInit, error value)
Error setting DAC
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (InitDacs, error value)
Error setting DAC
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (InstSettingInit, error value)
Error creating semaphore
2
(not significant)
Please return the unit to your local service center for repair
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
Tunable Laser Sources