E1x061: PGA DAC adj failed (a, b, c)
Tries to adjust the offset voltage of the PGA
a
shows the ADC offset reading
b
shows the limit for an upper offset
c
shows the limit for a lower offset
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x062: PGA gain adj failed (a, b, c)
Verifies the gain settings of the PGA (x 1/10/100). A voltage of
25 mV is applied to the input of the PGA and then amplified
a
if 1, the required division has failed
if 0, the calculated factors are correct
b
shows the measured values for an amplification by 10
c
shows the measured values for an amplification by 100
d
shows the quotient of b divided by c
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x063: Lambda DAC adj failed (a, b, c, d)
Verifies the gain of the responsiveness correction stage
a
if 1, the quotient d could not be calculated, otherwise 0
b
shows the difference between DAC settings 1024 and
2048
c
shows the difference between DAC settings 2048 and
4096
d
shows the quotient of b divided by c
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x080; TEC settling failed (a, b)
Verifies the comparator output of the regulation circuit after the
TEC is enabled and expects an ADC reading of about 512
a
shows the ADC reading of the comparator output
b
shows the time when the temperature regulation has
reached the default temperature. Waits maximum 30 sec-
onds
x
refers to the module channel tested
Please return the unit to your local service center for repair