Applications
234
Setting Analysis Parameters
22 To edit the power units used for the analysis, move to [W], press
Enter, move to <dBm>, <W>, or <dB>, and press Enter. See "What
are the Power Units ?" on page 80 for an explanation of power
units.
23 To edit the reference mode used for the analysis, move to
<Ref Mode>, press Enter, move to one of the following referencing
modes:
– <Value>, you set a reference value using <Ref>, see step 24,
– <First Sample>, the first sample of the logging function is
selected as the reference value,
– <Channel 1>, the value that Channel 1 measures is updated as the
reference value for each set of power measurements (thus, each
data value for channel 1 will be 0 dB), and
press Enter.
N O TE
These settings have an effect when you choose <dB> as the
<Pwr unit>. See step 8 or step 22.
24 If you have chosen <Value> as the <Ref Mode>, move to <Ref>, press
Enter, enter a value for the reference, and press Enter.
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
The Logging Application