Optical Switch modules
error message
description
test value
action
error message
description
test value
action
error message
description
note
test value
action
error message
description
test value
action
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
– Error spawning tTempCtrl!
– Error spawning tPowCtrl!
– Error checking firmware CRC!
Optical Switch modules
InitError() failed!
Initialization of internal error buffer failed.
Please return the unit to your local service center for repair
Invalid Caldata!
The calibration data loaded on the module is invalid.
Either the version number is nir compatible with the firmware or the
calibration data is corrupt (checksum incorrect).
Please return the unit to your local service center for repair
InitModule() failed!
Module initialization failed.
Normally something more descriptive appears (i.e. one of the other
messages mentioned here.)
Please return the unit to your local service center for repair
SanityChecks():channel X, configArray != -1
SanityChecks():channel X, configs == 0
SanityChecks() failed
Calibration data seems not to be correct
X Channel number
Please return the unit to your local service center for repair
Error Messages
403