The Logging Application
Agilent 8163A/B, 8164A/B, and 8166A/B User's Guide, Fourth Edition
N O T E
If you are using the Agilent 8163A/B Lightwave Multimeter, press
[Parameter] to access the Logging Parameter Setup Screen, as shown
in Figure 147. To rewturn to the Logging Modules Setup Screen, press
[Modules].
Figure 147
The Agilent 8163B Logging Parameter Setup Screen
7 Press [Menu] to access the Logging application menu screen.
8 Move to <Pwr unit>, press Enter, move to <dBm>, <W>, or <dB>, and
press Enter. See "What are the Power Units ?" on page 80 for an
explanation of power units.
9 Move to <λ>, press Enter, enter the wavelength value of your optical
source, and press Enter.
N O T E
Make sure that you install Power Meters that have similar
wavelength ranges. The wavelength range of the application is the
overlapping wavelength range of all installed Power Meters
10 Move to <Range Mode>, press Enter, move to one of the following
range modes:
– <common>, you set the same <Range> for each Power Meter,
– <individual>, you must choose an individual <Range> for each
Power Meter, and
press Enter.