Performance Test; Equipment - Agilent Technologies ESA Series User's And Service Manual

Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

E440xBU Option H70

Performance Test

Equipment

Gather the equipment listed in the following tables to verify the performance of the ESA
amplitude at the 70 MHz IF Out on the rear panel.
Table 3
Equipment List
Description
Signal Source Equipment
Synthesized Sweeper
Analyzer Equipment
Spectrum Analyzer
Power Meter Equipment
Power Meter
Power Sensor
Cables and Attenuators (as needed)
APC 3.5 (m)(m)
2 required
SMA
Installation Guide
Characteristics
Frequency: Must sweep 100 MHz
around 1 GHz Amplitude: −10 dBm
Upper Frequency Range: 2 GHz
Absolute Accuracy: ±0.5%
Resolution: 0.01 dB
Reference Accuracy: 1.2%Power
Reference Accuracy: ±0.9%
Compatible with Agilent 8480 series
power sensors dB relative mode
Frequency Range: 1 MHz to 4.0 GHz
SWR at 50 MHz: ≤ 1.05
1 MHz to 4 GHz SWR: ≤ 1.22:1
Type-N (m), 50 Ω
DC to 26.5 GHz
Length: ≤ 92 cm (36 in)
Insertion Loss: ~2 dB
Low-loss, 1 meter
Performance Test
Model
83620B
8562E
8563E
E4419B
E4418A
8482A
8481A
8120-4921
5064-5458