SET UP

CONTROL SET UP

6.2
CONTROL SET UP

CBC/DIFF file

1.
Select
Control set up
2.
Select a file to set up.
3.
Insert the 5C cell control diskette into
the diskette drive of the computer.
4.
Press
5.
Press the function key for the desired
level of control.
É for Normal
Ê for Abnormal I
Ë for Abnormal II
6.
Manually enter Shift and Operator ID.
7.
Check that HOST: is set according to
your laboratory protocol. ON means
that control run results are transmitted
to your host computer at the time of the
run. ß toggles between ON and
OFF .
8.
Check all entries to make sure they are
correct then press Ò to save and
escape.
9.
Repeat steps 2 through 8 for the other
levels of control. Once you are finished,
remove the 5C cell control diskette from
the diskette drive of the computer.
6-2
Special Functions
Set Up
tt
.
CBC/DIFF file
tt
Í Upload Assay Values
.
tt
PLACE 5C CONTROL DISK IN DRIVE A
Select Control Level:
F1: Normal
F2: Abnormal I
F3: Abnormal II
PN 4237519CB