SET UP
SAMPLE ANALYSIS SET UP
6.3
SAMPLE ANALYSIS SET UP
Action limits
Here you set the limits for XB, Definitive flags and High/Low flags according to your
laboratory protocol. Enter your Laboratory Normal Ranges in the DMS if you want them to
print on the report. Access to this area of the software requires the supervisor password.
For more information on XB, refer to Chapter 2. For more information on flags, refer to
chapter 4.
Note: If flagging limits are changed, flag assignments on previously stored data will not
update and will not correspond to the new limits.
XB limits
Enter XB target values and thier limits here.
1.
Select
Sample analysis set up
XB limits
2.
Move the cursor to the desired field.
Type the data then press Û.
3.
4.
Repeat steps 2 and 3 until all data has
been entered.
Press Ò to save and escape.
5.
Definitive flag limits
Enter your definitive flag limits here. As you
highlight each limit, the flag associated with
it appears at the top of the screen. If a
particular limit does not have a flag
associated with it, the area at the top of the
screen will be blank.
1.
Select
Sample analysis set up
Definitive flag limits
2.
Move the cursor to the desired field.
Type the numeric limit then press Û.
3.
4.
Repeat steps 2 and 3 until all data has
been entered.
Press Ò to save and escape.
5.
6-6
Special Functions
Set Up
tt
Action limits
tt
.
Special Functions
Set Up
tt
Action limits
tt
.
tt
tt
tt
tt
PN 4237519CB