Advertisement

5.
Mix according to the tube
manufacturer' s instructions.
6.
Identify the sample:
Hold the tube' s bar-code label in
r
front of the reader.
Green light and beep:
Bar code read. Go to step 7.
Red light:
Wait. Try again.
Note: Cycle the sample within 10
seconds of reading the bar code.
After 10 seconds, the system
deletes the identification.
Enter 1 to 16 alphanumeric
r
characters then press Û.
7.
Cycle the sample:
a.
Open the tube and immerse the
aspirator tip
b.
Press and release the sample bar.
c.
Remove the tube when you hear
the beep.
8.
Review the results. Refer to Chapter 4,
Data Review, for information on the
Run Samples screen, scatterplots,
histograms, parameter codes and flags
and messages.
PN 4237519CB
OR
b
into the sample.
CYCLING SAMPLES IN THE SECONDARY MODE
1
3
SAMPLE ANALYSIS
3-9