3.4

TUBE ADAPTERS

Round-bottom tube adapters can be used for:
2mL (10.25mm X 47 mm) or
3mL (10.25 mm X 64 mm) sample tubes.
You can use a maximum of two labels in
r
addition to the sample tube
manufacturer' s label.
Place each label so that it does not cover
r
the bottom of the tube and is flat and
smooth against the tube. This prevents
the adapter from being broken or
jammed.
Ensure that the bar-code symbol and a
r
1/4-inch blank space "quiet zone" on
either side of the bar-code symbol are
visible through the read slit
you insert the tube into the adapter.
If the entire bar-code and "quiet zones"
r
do not fit within the read slit, the
number of data digits that can be read
from your bar-code label may be
reduced. In this case, read the bar-code
label before you put the tube in the
adapter then cycle immediately.
3.5

PREASSIGNING THE WORKLIST

If using this option, assign samples to the
Worklist now.
See Heading 3.11, Worklist at the end of this
chapter for more information and
procedures.
3.6

SAMPLE INTEGRITY CHECKS

In Primary mode, the system checks each sample aspiration using dual sensors, called blood
detectors, which monitor the blood before and after it passes through the Blood Sampling
PN 4237519CB
b
when
2 mL
SAMPLE ANALYSIS
TUBE ADAPTERS
3 mL
3-3
3