10. Cycle the control (bottle 2):
IMPORTANT
Do not remove the control bottle until you hear the beep.
a.
Immerse the aspirator tip
completely in the control.
b.
Press and release the sample bar.
c.
Remove the control bottle when
you hear the beep.
11. Check for H (High) or L (Low) beside
the results for both modes.
If there are no H's or L's, results are
r
within range.
If you see an H or L, go to
r
Table 2.1. Follow the
troubleshooting steps until you
solve the problem.
12. Optional: Press Ì to print the results
for your logbook.
PN 4237519CB
Removing the control bottle before you hear the beep can cause misleading control results.
STARTUP AND CONTROLS
CONTROL RUN
2