Beckman Coulter COULTER HmX Operator's Manual page 63

Hematology analyzer
Hide thumbs Also See for COULTER HmX:
Table of Contents

Advertisement

IMPORTANT
foaming or bubbles occur, but rapidly enough to mix the blood-stain mixture and Reagent B. Do not do
any additional mixing.
5.
Dispense Reagent B.
a.
Place tube "B" with the blood-stain
aliquot at a 30° angle under the tip
of the Reagent B dispenser.
b.
Dispense 2 mL of Reagent B into
the test tube "B." DO NOT MIX.
6.
Wait 30 seconds.
While you wait, enter the sample' s
identification. Type 1-16 alphanumeric
characters then press Û.
7.
After 30 seconds, analyze the sample.
a.
Immerse the aspirator tip
the retic preparation.
b.
Press and release the sample bar.
c.
Remove the tube when you hear
the beep.
PN 4237519CB
To ensure accurate results, allow Reagent B to run down the side of the tube so that no
b
CYCLING SAMPLES IN THE RETIC MODE
into
1
3
SAMPLE ANALYSIS
2 mL
3-17