1.5

RUN SAMPLES SCREEN OPTIONS

Í

Optns:

F2
XB:
F3
CAP PIERCER 10mm-13mm
F4
DB:
F5
Print:
F6
Host:
F7
Display only: OFF
F8
Operator: OPR
F11
B&W screen print
F12
Color screen print
Table 1.1 F5 from Run Samples Screen
F2 XB:
F3 CAP PIERCER 10mm-13mm Alternate between this and 16mm. Select the correct carousel slot
F4 DB:
F5 Print: NONE
F6 Host:
F7 Display only: OFF
F8 Operator: OPR
F11 B&W screen print
F12 Color screen print
Note: After a system reset, these options return to their default settings. Be sure to set them
up again according to your laboratory' s protocol before running patient samples.
Note: Print, Host and Operator can also be set up from the Main Menu using
PN 4237519CB
ON
N=2 IN
ON
NONE
OFF
ON N=2
IN
Turns XB ON and OFF. N is the number of samples stored in the current
batch. Also displays the status of the last completed batch (IN or OUT).
based on the size of the tube you run.
ON
Turns the data base ON and OFF. Default setting is ON.
Sets the automatic printing of samples to the graphic printer. Choose
between NONE, NORMALS, ABNORMALS or ALL. Default setting is
NONE.
OFF
Turns the automatic host transmission ON and OFF. Default setting is
OFF.
If ON, then XB, DB and HOST turn OFF. Default setting is OFF.
Enter up to three alphanumeric characters for an Operator ID. Default
setting is OPR.
Initiates a large black and white screen print of the current sample.
Initiates a large color screen print of the current sample if your printer
can print in color.
SYSTEM OVERVIEW
RUN SAMPLES SCREEN OPTIONS
F5-Options