9.
Place the control in the carousel.
Results are placed automatically in
r
the correct file.
Results do not appear on the Run
r
Samples screen.
If any result is out of control, an
r
error message displays.
10. Repeat steps 7 through 9 for other levels
of control.
11. Check the results of the controls.
a.
Select
b.
Check for H (High) or L (Low)
beside the results.
Note: For more information about
Review or Report,
Use Ê
c.
review.
12. Optional: To print your last control run,
select
. Use Ê
Print
print.
Table 2.2 When CBC/DIFF Control is Out of Limits
Possibility
Improper mixing
Control file set up incorrectly
Chance (statistical outlier)
Change in the control
Instrument change
PN 4237519CB
Controls tt Review or Report.
If there are no H's or L's,
r
results are within range.
If you see an H or L, go to
r
Table 2.2. Follow the
troubleshooting steps until
you solve the problem.
see Heading 2.3
File
to select other files to
. Press Ì
Controls tt Control Run
File
to select other files to
STARTUP AND CONTROLS
Action
Follow the instructions on the package insert. Rerun
control.
Make sure the assigned values and ranges match
those on the control package insert. If in error,
correct them by selecting Special Functions tt Set
Up tt Control Set Up.
Rerun the control. If it is still "out," try the next
possibility.
Try another vial or level of control. Follow directions
on the package insert for proper handling.
Watch for normal sample flow. Call your Beckman
Coulter Representative to help you troubleshoot
abnormal operation.
2
CONTROL RUN
2-9