STARTUP AND CONTROLS
CONTROL RUN
IMPORTANT
tube; do not allow control-stain mixture to run down the sides of the tube. To prevent drying of this
small amount, proceed immediately to the next step.
10. Gently mix tube "A" again then transfer
2 µL of the control/stain mixture from
tube "A" into the bottom of tube "B."
IMPORTANT
foaming or bubbles occur, but rapidly enough to mix the control-stain mixture and Reagent B. Do not do
any additional mixing.
11. Dispense Reagent B.
a.
Place tube "B" with the
control-stain aliquot at a 30° angle
under the tip of the Reagent B
dispenser.
b.
Dispense 2 mL of Reagent B into
the test tube "B." DO NOT MIX.
12. Wait 30 seconds.
2-16
To ensure accurate results, add the control-stain mixture directly to the bottom of the
To ensure accurate results, allow Reagent B to run down the side of the tube so that no
2 µL
2 mL
PN 4237519CB