Beckman Coulter COULTER HmX Operator's Manual page 22

Hematology analyzer
Hide thumbs Also See for COULTER HmX:
Table of Contents

Advertisement

STARTUP AND CONTROLS
CONTROL RUN
6.
Evaluate primer results:
a.
Are both counts < 500?
b.
Cycle a new vial. Make sure it is
free of bubbles. Are both counts
Press È to remove the Primer Run
7.
window.
8.
When SELECT FUNCTION appears on
the status line, press Ë
Ë
CONTROL (SECONDARY)
9.
Gently mix the control. Invert the bottle
5 to 8 times.
2-4
If yes, go to step 7.
r
If no, go to step 6b.
r
< 500?
If yes, go to step 7.
r
If no, press Ì to print the
r
screen then call your Beckman
Coulter representative.
Run
.
SAMPLE MODE?
F3
CONTROL (SECONDARY)
F4
PRIMER
F7
PURGE
F8
RINSE
F9
STOP
Select to change/ESC to continue
PN 4237519CB