High/low flag limits

Enter High/Low flag limits here. The H or L
will appear next to the numeric result that
exceeds the limit.
1.
Select
Sample analysis set up
High/low flag limits
2.
Move the cursor to the desired field.
Type the numeric limit then press Û.
3.
4.
Repeat steps 2 and 3 until all data has
been entered.
Press Ò to save and escape.
5.
Note: MRV and IRF parameters only appear
on the screen if you enabled these
parameters.

Laboratory Normal Ranges

Enter your laboratory' s normal ranges here.
These do not trigger any flags. They serve
only as a reference on printed reports.
1.
Select
Sample analysis set up
Laboratory Normal Ranges
2.
Move the cursor to the desired field.
Type the data then press Û.
3.
4.
Repeat steps 2 and 3 until all data has
been entered.
Press Ò to save and escape.
5.
Note: MRV and IRF parameters only appear
on the screen if you enabled these
parameters.