Cycling COULTER 5C Cell Controls in Primary Mode with Bar-Code Labels

COULTER 5C cell control, with bar-code labels, is the recommended method of QC. Your
HmX Hematology Analyzer automatically recognizes bar code-labeled 5C cell controls and
assigns results to the correct file. If the bar-code label cannot be read, follow the procedure
Cycling Commercial Cell Controls without Bar-Code Labels.
IMPORTANT
1. If you cycle 5C cell control with the DIFF OFF, differential results do not post to the control file and
therefore are not evaluated for being IN or OUT of control. Cycle 5C cell control with the DIFF ON.
2. Misleading results can occur if 5C cell control is not prepared properly. Follow the procedure on the
package insert to properly warm and mix 5C cell control.
1.
Follow the directions on the cell control
package insert for storage, preparation
and mixing.
2.
Does SELECT FUNCTION appear at the
lower right corner of the DMS screen?
If no, go to step 3.
r
If yes, continue with this step.
r
a.
Access the Run Samples screen:
b.
The instrument automatically
prepares itself for Primary mode,
DIFF ON. Go to step 7.
3.
Access the Run Samples screen:
at the Access screen,
r
press É
at the Main Menu, select
r
Analysis tt Run Samples
Press Ë
4.
PN 4237519CB
at the Access screen,
r
press É
RUN SAMPLES
OR
at the Main Menu, select
r
Sample Analysis tt Run Samples
RUN SAMPLES
OR
Run
.
.
Sample
STARTUP AND CONTROLS
CONTROL RUN
2-7