Edit A Sample; Results Of Changed Parameters - Beckman Coulter COULTER HmX Operator's Manual

Hematology analyzer
Hide thumbs Also See for COULTER HmX:
Table of Contents

Advertisement

DATA REVIEW
DATA BASE QUERY

Edit a Sample

1.
Select
2.
Perform a sort that includes the sample
you want to edit.
3.
Highlight the patient sample to edit.
Press Ë
4.
Results screen.
5.
Move to fields that need to be edited
with æ ç è and é keys.
Highlight the field and type the new
data. If the date of birth is prior to 1990,
enter all four digits of the year.
6.
An E appears in all edited results fields.
7.
When you edit a non-cycled parameter,
the system backlights the result field.
When you edit either the RBC result or
RET% result, the system computes
RET# and displays an e next to RET#,
indicating that it is computed using an
edited parameter. The system also
displays an H or L if RET# is higher or
lower than the defined limits.
Press Ò
8.

Results of Changed Parameters

If you change any parameter result:
All population Suspect and Definitive messages are deleted from the data base.
r
The message EDITED DATA appears in the Population message field.
r
No Population Suspect or Definitive messages are printed or transmitted to a host
r
computer.
The EDITED DATA message appears on all printouts.
r
An E and/or e appears on all printouts next to the edited result.
r
Data transmitted to a host computer includes the EDITED DATA message and the E/e
r
flags.
4-20
Sample Analysis tt Data Base Query.
to go to the Sample
Edit
Save/Esc
.
PN 4237519CB