Calibrate the printer
Calibrate the printer when you see the calibration indicator
dashboard. (See
IMPORTANT
You are strongly advised to perform a daily printer calibration.
Tools
Eye-One spectrophotometer
Before you begin
Media for calibration
Prepare calibration media according to the calibration media specifications. (See
calibration on page
Procedure
1.
Load the calibration media.
NOTE
Although it is advised to perform the default selected calibration procedures, you can touch
[Printer calibration details] to change the following selections:
• [Shading correction].
• [Auto gradation adjustment]
2.
Connect the Eye-One spectrophotometer to the USB port in the control panel.
3.
Touch [System] -> [Color adjustment] -> [Start printer calibration].
4.
Follow the instructions on the operator panel.
5.
Select [Automatic (internal sensor)] in the [Calibration method] window.
About calibration on page
354)
on the right-hand side of the
353)
Chapter 14 - Maintenance and print quality
Calibrate the printer
Media for
355