Adjusting the detection point of the touch screen
(Calibration function)
Follow the steps below to correctly align the positions of the LCD display and the touch screen.
1.
While holding down the SCENE MEMORY [STORE] key on the panel, turn on the
power to the CL unit.
After the opening screen, the following startup menu screen will appear.
2.
Press the TOUCH SCREEN CALIBRATION button.
The TOUCH SCREEN CALIBRATION MODE screen will appear, allowing you to calibrate the
touch screen.
NOTE
If you cannot start calibration by pressing the button, press the SCENE MEMORY [INC]/[DEC]
key to select TOUCH SCREEN CALIBRATION, and then press the [STORE] key to start.
3.
Press the START button.
A confirmation dialog box will appear.
4.
Press the OK button in the dialog box.
A cross-shaped cursor will appear in the screen.
5.
This cursor will appear a total of three times. Press each location of the cursor at
which it appears.
NOTE
To set the detection points accurately, press the cross-shaped cursor from the position and
posture in which you normally operate the unit.
6.
Press the EXIT button.
The CL unit will start up in normal operating mode.
NOTE
Alternatively, you can continue operation by selecting a different menu instead of pressing the
EXIT button.