2.3.6 Installations Requirements & Environment Control
64
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
NOTICE
In order to minimize interference risks, the following requirements shall apply:
Cables shielding & grounding:
All interconnect cables to peripheral devices must be shielded and properly
grounded. Use of cables not properly shielded and grounded may result in the
equipment causing radio frequency interference.
Separated Power supply distribution panel & separated power line:
This product complies with the radiated emission limits as per the CISPR11
Group1 ClassA standard at the premises of a typical user. .
The Innova system is predominantly intended for use (e.g. in hospitals) with
a dedicated supply system, and with an X–ray shielded room.
In case of using in a domestic environment (e.g. doctors' offices), in order
to avoid interferences, it is recommended to use a separated AC power
distribution panel & separated power line, and an X–ray shielded room.
Subsystems, options & accessories Power supply distribution:
All components, accessories subsystems, systems which are electrically
connected to the Innova system, have to be all AC power supplied by the same
power distribution panel & line.
NOTE:
In order to avoid interferences, the same AC power distribution panel
should supply all Innova system components , accessories,
subsystems and options as the Advantage Workstation, and the
UPS. That same AC power distribution panel should be supplied by
a separated AC power line (coming from a separated transformer line
and winding).
Stacked components & equipment:
The Innova system should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the Innova system should
be observed to verify normal operation in the configuration in which it will be
used.
Low frequency magnetic field:
In case of an Innova system, the Gantry (digital detector) shall be apart 1meter
from the generator cabinet, 1m. from the PDB cabinet, 1 m. from the UPS
cabinet, and 1 meter apart from monitors. These distances specifications shall
minimize the low frequency magnetic field interference risk.
Static Magnetic Field Limits : The Innova system is compatible with the Earth
magnetic Field. Earth magnetic Field is lower than 1 Gauss.
Electrostatic discharges environment & recommendations:
In order to reduce electrostatic discharge interference, install a charge
dissipative floor material to avoid electrostatic charge buildup.
The relative humidity shall be at least 30 percent.
The dissipative material shall be connected to the system ground reference,
if applicable.
Direction 5177951-4-1EN, Revision 1