GE Innova 2121-IQ Biplane Preinstallation Manual page 111

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
Maximum cable drape length is 6 m / 236 in (with vinyl zipper cable cover of 6.2 m / 244 in) from
LP4 to ceiling exit point. This includes sag between drape points (A, B, C and D).
The worst case LP4 cable drape extension, including sag, is:
1.85 m (72.8 in) between A and B
2.05 m (80.7 in) between B and C
2.30 m (90.5 in) between C and D
Chapter 5 Physical Characteristics
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 5-13: LP4 Positioner Cable Drape Length
Direction 5177951-4-1EN, Revision 1
111