GE Healthcare

2.4 Injector Mounting Requirements

144
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
CAUTION
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg at 150 mm (60 N.m). Therefore:
Only light extra load not exceeding 5 kg at 100 mm (i.e IV pole with its
accessories, pressure head...) is authorized on the same table accessory rail
as the injector.
Never install injector and radiation protection on the same table accessory rail.
Typical installation on the front table accessory rail is Smart handle or Smart
box, Table Side System Control (TSSC), InnovaCentral/Touchscreen, Table
panning device and cables support.
If needed an optional rail can be installed at table foot end of the Omega V table
for other options.
Direction 5177951-4-1EN, Revision 1
2 Mounting Requirements