GE Healthcare

5 IVUS Rev 1 Option

5.1 Wiring Diagram

The connection cables between the IVUS Control Room and Procedures Room components run
in a dedicated under floor conduit (see
A GND wire between the C1 cabinet and the IVUS CPU runs in the underfloor conduit prepared
for the Control Booth cables between Equipment Room and Control Room.

5.2 IVUS Floor Conduits

A 3 inches and max. 25 m long conduit between Control and Patient Rooms shall be prepared
below the floor for the IVUS Innova interconnection cables.
Locate the position of the IVUS components in the Control Room, according to the workflow
requirement of the operators.
Chapter 6 Electrical Connections
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Section
5.2).
Illustration 6-10:
Direction 5177951-4-1EN, Revision 1
159