PRODUCT OR
COMPONENT
UPS 3 kVA - model
9130
Power Distribution
Box CE
Power Distribution
Box UL
DL keypad
DL LCD monitor
VCIM
Videostation VCR
See note 3
6 monitors suspen‐
sion
8 monitors suspen‐
sion
Components common to IVUS Rev 1 & Rev 2 Option
IVUS PIM (Ultra‐
sound probe)
IVUS Control Sta‐
tion
IVUS Joystick
IVUS Isolation
transformer
IVUS CPU
IVUS Keyb. &
mouse pad
IVUS monitor
IVUS Image printer 6 (13.2)
IVUS 4-way Video
Switch
Components specific to IVUS Rev 2 Option
PIMr
Pimette
Touchpad controller 1 (2,2)
Printer HP
NOTE:
(1) including 353 lbs (160 kg) patient.
NOTE:
(2) Depth.
NOTE:
(3) Maximum dimensions given. Exact dimensions depend on chiller manufacturer type.
Chapter 5 Physical Characteristics
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
NET
DIMENSIONS MM (INCHES)
WEIGHT
WIDTH
DEPTH
KG (LBS)
36 (79)
214 (8.43) 412 (16.2)
300 (11.8) 1800
225 (496) 800 (31.5)
389.5
356 (14)
1016 (40)
(859)
283
1.4 (3)
300 (11.8) 82 (3.25)
(11.55)
8.2 (18)
179 (7)
387 (15.2) 504 (19.8)
0.95
450 (17.7) 150 (5.9)
(2.09)
370
270
12 (26.5)
(14.57)
(10.63)
126 (277)
170 (375)
2 (4.4)
38
96
4 (8.8)
120
360
1 (2.2)
90
76
7 (15.4)
97
267
15 (33.1)
396
162
1 (2.2)
NA
NA
15 (33.1)
420
432
158
184
1 (2.2)
44
222
1,5 (3,2)
356
102
1 (2,2)
142
76
211
180
1,3 (2,8)
227
138
LOAD BEARING
AREA MM (IN‐
HEIGHT
CHES)
346 (13.6)
with feet
(70.8)
2120 (83)
50 (2)
144 (5.67)
177
240
102
171
510
NA
240
18
241
102
18
76
117
Direction 5177951-4-1EN, Revision 1
WEIGHT/ OC‐
MOUNTING METHOD
CUPIED AREA
125