Table 3-3:
Room
Core System
Exam
Innova LC/LP positioner
Room
and table
Ctrl Room DL user area with 1 TFT
monitor
3 B&W flat monitors
Ctrl Room Total
Tech
C1 Frontal Cabinet
Room
C1 Lateral Cabinet
C2 Cabinet
Coolix X-Ray tube chiller
Frontal (1) (2)
Coolix X-Ray tube chiller
Lateral (1) (2)
Detector conditioner
Frontal
Detector conditioner Lat‐
eral
Main disconnect panel
PDB
3kVA Cabinet UPS - mod‐
el 9125
3kVA Cabinet UPS - mod‐
el 9130 (4)
Tech Room Total
Total for core system
Room
Options (3)
7 in room B&W TFT mon‐
itors
Exam
In room AW TFT monitor 0.12
Room
Typical injector
Exam room total
AW work station
2 AW TFT monitors
Ctrl Room
Printer
Ctrl room total
Typical configuration without fluoro
UPS
Note (1): Air flow requirements 1200 m 3 /h (706 CFM)
Note (2): For more details, consult appropriate pre–installation manual
Chapter 3 Physical Requirements of room
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
HEAT OUTPUT
Stand by
kW
0.61
0.16
0.25
0.41
0.71
0.31
0.29
2.53
2.53
0.21
0.21
0.60
0.16
0.37
7.54
8.56
Stand by
0.59
0.09
0.80
0.35
0.24
0.31
0.90
10.27
Moderate Use (8 cas‐
es / a 10 hour day)
BTU/hr
kW
BTU/hr
2076
0.75
2540
546
0.16
546
859
0.25
859
1405
0.41
1405
2421
0.99
3389
1057
0.56
1923
989
0.83
2813
8619
4.49
15309
8619
4.19
14274
709
0.21
709
709
0.21
709
2216
0.65
2216
546
0.16
546
1257
0.37
1257
25885
12.29
41888
29366
13.44
45832
Moderate Use
2005
409
320
Same values as Stand
2734
by
1201
818
1054
3073
35173
15.15
51640
Direction 5177951-4-1EN, Revision 1
Maximum Use (maxi‐
Typical Use (13 cas‐
mum peak power dur‐
es / a 10 hour day)
ing exam)
kW
BTU/hr
kW
1.21
4128
1.62
0.16
546
0.16
0.25
859
0.25
0.41
1405
0.41
1.29
4412
1.59
0.86
2946
1.16
1.34
4571
1.81
5.49
18725
6.93
4.79
16319
6.93
0.21
709
0.21
0.21
709
0.21
0.65
2216
0.65
0.16
546
0.16
0.37
1257
0.37
15
51153
19.65
16.63
56685
21.68
Typical Use
Maximum Use
Same values as
Same values as Stand
Stand by
by
18.33
62493
23.39
BTU/hr
5517
546
859
1405
5435
3969
6171
23625
23625
709
709
2216
546
1257
67003
73925
79732
57