2 Room Layout Drawings

Table 4-3: Exam room dimensions
Room Dimensions
Recommended:
Minimum:
Table 4-4: Room Layout components (see
Chapter 4 Room Layouts
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS
MANDATORY. THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER,
DETECTOR CHILLER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS
THAT ARE CREATING AIR–CIRCULATION OF PULSED–AIR. WHEN THIS
PULSED AIR IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE
PATHOGENS LIKE AN EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF
TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM PATIENTS TO
OTHER PATIENTS OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED
IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E.,
TECHNICAL ROOM.
Length x Width
11570 mm x 8000 mm (38 ft 0 in x 26 ft 2 in)
6900 mm x 4400 mm (22 ft 8 in x 14 ft 5 in)
1
2
PDB
3
C1F
4
C2
5
C1L
6
F-UPS
7
Det Cond LAT
8
Chiller LAT
9
Det Cond FRT
10
Chiller FRT
11
12
13
LP4
14
LC
15
16
17
TBL
18
19
20
WBC1
21
Direction 5177951-4-1EN, Revision 1
Ceiling Height
2845 mm ± 5 mm (9 ft 4 in ± 0.2 in) is mandatory
ceiling height
Illustration
4-3):
3 kVA UPS cabinet
Power Distribution Panel
C1 Frontal Cabinet
Positioner cabinet
C1 Lateral cabinet
Fluoro UPS
Lateral Detector conditioner
Lateral Tube chiller
Frontal Detector conditioner
Frontal Tube Chiller
LP4 stationary rails
Cable drape rail
Lateral Gantry
Frontal Gantry
6 or 8 LCD Monitor suspension
Counter balanced eye or thyroid shield
Omega table
Customer supplied storage cabinet
XT Stationary Rails
Operators console
AW workstation
91