Environmental Requirements/Limitations; Room Climate - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare

2 Environmental Requirements/Limitations

2.1 Room Climate

Innova system room climate requirements – relative humidity and temperature (individual products
or components are classed by their installation area)
Table 3-1:
INSTALLATION
ROOM OF
PRODUCT OR
COMPONENT
Examination
room
Technical room
(See Note (1))
Technical room
with Fluoro UPS
optional
Control room
Innova Digital
Detector
NOTE:
(1): The target temperature (best recommended) is 18°C (64°F).
The target humidity (best recommended) is 50% RH.
54
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
NOTICE
Avoid extremes in temperatures
RELATIVE HUMIDITY (NON- CONDENS‐
ING)
IN-USE
MIN
MAX
MIN
30%
75%
10%
30%
75%
10%
30%
75%
10%
30%
75%
10%
See Note (2)
IN-USE (See Note
STORAGE
(3a))
MAX
MIN
+15°C
70%
+59°F
+10°C
70%
+50°F
+20°C
70%
+68 °F
+10°C
70%
+50°F
Direction 5177951-4-1EN, Revision 1
TEMPERATURE
RECOMMENDED
(See Note (3b))
MAX
MIN
MAX
+32°C
Design for Patient/
+90°F
Operator Comfort
+35°C
+13°C
+25°C
+95°F
+55°F
+77°F
+25°C
+20°C
+25°C
+77°F
+68 °F
+77°F
+35°C
+20°C
+25°C
+95°F
+68°F
+77°F
2 Environmental Requirements/Limitations
STORAGE / TRANS‐
PORT
MIN
MAX
-10°C
+50°C
+14°F
+122°F
-14°C
+50°C
+8°F
+122°F
-5°C
+50°C
+23°F
+122°F
-10°C
+50°C
+14°F
+122°F