GE Healthcare
NOTE:
(4) For regions without seismic constraints, the Detector Conditioner Thermo-Con can
be floor or wall mounted. For regions with seismic constraints, the Detector Conditioner
Thermo-Con must be floor mounted.
2.2 Positioner and Table Floor Mounting
The distances between the Innova LC Positioner and the Tables are critical for a proper clinical
usage. For this reason, GE Healthcare provides two floor mounting templates to ensure these
components are properly placed in relation to one another.
Table 5-4:
Innova LC Positioner Floor Mounting Methods
Cable Conduit For On-Grade Floor Anchor Kit
Inner Base Plate For Above Grade Floor Anchor Kit
Fixing Bolt Overview
Gantry and table mounting holes
126
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Title
Direction 5177951-4-1EN, Revision 1
Illustration
Illustration 5-26
tion 5-27
Illustration 5-28
Illustration 5-29
Illustration 5-30
Illustration 5-31
2 Mounting Requirements