Installations Requirements & Environment Control - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2.4.6 Installations Requirements & Environment Control
64
In order to minimize interference risks, the following requirements shall apply:
Cables shielding & grounding:
All interconnect cables to peripheral devices must be shielded and properly
grounded. Use of cables not properly shielded and grounded may result in the
equipment causing radio frequency interference.
Separated Power supply distribution panel & separated power line:
This product complies with the radiated emission limits as per the CISPR11
Group1 ClassA standard.
The Innova system is predominantly intended for use (e.g. in hospitals) with
a dedicated supply system, and with an X–ray shielded room.
In case of using in a domestic environment (e.g. doctors' offices), in order
to avoid interferences, it is recommended to use a separated AC power
distribution panel & separated power line, and an X–ray shielded room.
Subsystem & accessories Power supply distribution:
All components, accessories subsystems, systems which are electrically
connected to the Innova system, have to be all AC power supplied by the same
power distribution panel & line.
NOTE:
In order to avoid interferences, the same AC power distribution
panel should supply all components, accessories, the Innova
system (& subsystems as the Advantage Workstation). The
separated AC power line should supply the panel.
Stacked components & equipment:
The Innova system should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the Innova system should
be observed to verify normal operation in the configuration in which it will be
used.
Low frequency magnetic field:
In case of an Innova system, the Gantry (digital detector) shall be apart 1 meter
from the generator cabinet, and 1 meter apart from the analog (CRT) monitors.
These distances specifications shall minimize the low frequency magnetic field
interference risk.
Static magnetic field limits:
In order to avoid interference on the Innova system, static field limits from
the surrounding environment are specified.
Static field is specified less than <1 Gauss in Examination room, and in the
Control Area.
Static field is specified less than <3 Gauss in the Technical Room.
Electrostatic discharges environment & recommendations:
In order to reduce electrostatic discharge interference, install a charge
dissipative floor material to avoid electrostatic charge buildup.
The relative humidity shall be at least 30 percent.
The dissipative material shall be connected to the system ground reference,
if applicable.
Direction 5160944-13-1EN, Revision 2