GE Healthcare
18 Physical Runs
Suspension are always pre–cabled with 36 meter cables of connector to connector Monitor options
includes their 24 meters length for cabling
In case 24 m required, they have to be added in the order and replaced on site.
For a description of how to use the following 5 cable group schematics, see below:
Cable length data is as follows:
•
Cable Length = the total cable length, connector to connector (example above is 24 meters).
•
X + Y = used length for connection within system (example above is 5 meters).
•
Cable Length - (X + Y) = available length for conduit run (example above is 19 meters).
180
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
NOTICE
All lengths of cable are:
•
in useable meter when you look at group level, or
•
in meters (connector to connector) when you look at the cable level.
Illustration 6-19: Description of cable group diagrams
Direction 5177951-4-1EN, Revision 1
18 Physical Runs