Structural Requirements; Room Size; Elevator And Door Size Requirements; Door Height - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare

3 Structural Requirements

3.1 Room Size

For Room size dimensions, refer to
For additional details, refer to

3.2 Elevator and Door Size Requirements

Minimum door sizes also apply to hallways and elevators. For additional details, refer to
7, Shipping Information

3.2.1 Door Height

The minimum door height (to accommodate Innova positioner on its dolly) is 2001 mm (78.8 in).

3.2.2 Door Width

The minimum door width needed (to accommodate the Innova LC/LP shipping dolly) is 1040 mm
(41 in) with the cable inlet and the dolly stabilizers removed.
NOTE:
Door widths are based on a
corridor. Calculations need to be made for accommodation of equipment through
narrower corridors.

3.2.3 Elevator Depth

The minimum elevator depth needed to accommodate the Innova LP4 shipping dolly is 2954 mm
(116.3 in).

3.3 Floor

3.3.1 General Vascular GE Healthcare Policy

GE Healthcare's Customer is responsible for the structural analysis and mounting of the base
plates. If GE Healthcare is forced to mount the base plate, the Local Customer Team must hire a
structural engineer to design and approve the mounting method and provide GE Healthcare with
an engineering report.
The floor level cannot exceed a general levelness of 5 mm (0.2 in) for any 2 meters (79 in).
66
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Chapter 4, Room Layout
NOTICE
The floor slabs on which the equipment is to be installed must have a levelness of
1 mm (0.04 in) per meter (40 in). Position of baseplates and table basement
depends on the type of installation. The two types of installation are given in
Illustration
3-2.
Chapter 4, Room Layout
Considerations.
straight–in approach requiring a 2440 mm (96 in) wide
Direction 5177951-4-1EN, Revision 1
Drawings.
3 Structural Requirements
Chapter

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3131-iq biplane

Table of Contents