Layout Constraints - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
NOTE:
The order of the cabinets above is C1 frontal, C1 lateral and C2 cabinets. This is just an
example, cabinets can follow different orders.

1.8 Layout Constraints

Table 4-1: System Layout constraints Technical and Exam room on same floor
Location
Technical room
Technical room
88
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
CAUTION
The service area dimensions shown above are minimum requirements. Service
areas must comply with local regulations if more stringent.
Constraint Name
Cabinets location
Tube chillers location
Illustration 4-2:
Parameter
Specification
Maximum relative
9 m (29 ft, 6 in)
distance
The chiller shall be
no more than 5 m (15
feet) below or 8 m
(25 feet) above the
X-ray tube
Direction 5177951-4-1EN, Revision 1
Comment
Make sure to provide enough
space for storing extra cable length
(e.g. Plinth)
1 Room Layout Considerations